Q.bloxx_A127
|
|
|
Q.bloxx Modul A127
- 4 voltage input channels
- 2 inputs for voltage measurement
measuring ranges ±40 V, ±120 V, ±400 V, ±1200 V
2 inputs for current measurement via shunt resistors measuring ranges ±80 mV, ±240 mV, ±800 mV, ±2400 mV
- Fast high accuracy digitalization
- 24 bit ADU,
50 kHz sample rate per channel with 4 channels,
100 kHz sample rate per channel with 2 channels
- Signal conditioning
- linearization, digital filter, average, scaling, min/max storage, RMS, alarm, power and efficiency calculation
- RS485 fieldbus interface
- up to 48 Mbps: LocalBus
up to 115.2 kbps: Modbus-RTU, ASCII
- Connectable to any Test Controller
- e.g. Q.station, Q.gate or Q.pac
- Galvanic isolation
- channel to channel to power supply and to interface isolation voltage 1200 VDC / 858 VACrms test voltage 5 kVDC over 1 minute
- Electromagnetic Compatibility
- according EN 61000-4 and EN 55011
- Power supply 10…30 VDC
- DIN rail mounting (EN 60715)
|
|
|